9–13 Sept 2024
Wanda Realm Beijing
Asia/Shanghai timezone
IBIC2024 Final Proceedings: https://accelconf.web.cern.ch/ibic2024/

High-resolution quad-channel picoammeter: characterization and commissioning

WEP64
11 Sept 2024, 14:20
1h 30m
China Hall 3

China Hall 3

Poster Presentation MC7: Data Acquisition and Processing Platforms WEP: Wednesday Poster Session

Speaker

Fernando Cardoso (Brazilian Synchrotron Light Laboratory)

Description

To address the high demand for precise low current measurements at the Sirius accelerator and its beamlines, a quad-channel high-resolution Ethernet picoammeter has been designed*. The instrument can measure currents ranging from picoampere to milliampere across eight selectable ranges, featuring integrated ADCs enabling sample rates of up to 2 ksps and synchronization capabilities. This work aims to describe the design, characterization, and calibration results of the instrument. Special attention will be given to evaluating trigger latency, synchronization outcomes, as well as the device’s installation and commissioning at beamlines, particularly for critical applications like on-the-fly scanning experiments. Furthermore, we will explore the interplay between trigger period, digital filter bandwidth, and front-end analog bandwidth to optimize signal-to-noise ratio in specific applications.

Footnotes

*M. M. Donatti, F. H. Cardoso, L. Y. Tanio, Development of a quad-channel high-resolution digital picoammeter for beam diagnostics. 2024 JINST 19 C03028. [DOI:10.1088/1748-0221/19/03/C03028]

I have read and accept the Privacy Policy Statement Yes

Primary author

Mr Lucas Tanio (Brazilian Synchrotron Light Laboratory)

Co-authors

Fernando Cardoso (Brazilian Synchrotron Light Laboratory) Mr Mauricio Donatti (Brazilian Synchrotron Light Laboratory)

Presentation materials