Qualitative measurements of bunch length at CLARA using coherent transition radiation

THPL174
11 May 2023, 16:30
2h
Sala Laguna

Sala Laguna

Poster Presentation MC6.T03: Beam Diagnostics and Instrumentation Thursday Poster Session

Speaker

Storm Mathisen (Science and Technology Facilities Council)

Description

Bunch length is an important metric for user experiments at the Compact Linear Accelerator for Research and Applications (CLARA). A prototype Bunch Compression Monitor (BCM) based on Coherent Transition Ration (CTR) was recently installed and commissioned to support recent user experiments. The intensity of CTR is measured using a pyroelectric detector. A noise cancellation scheme based on a second detector offset from the focus of the CTR was used to reduce the noise caused by the broadband nature of pyroelectric detectors. Qualitative measurements of the bunch length as a function of RF phase are presented, along with an overview of the system design. Plans for an improved system are also presented.

I have read and accept the Privacy Policy Statement Yes

Primary author

Storm Mathisen (Science and Technology Facilities Council)

Co-authors

Thomas Pacey (Science and Technology Facilities Council) James Jones (Science and Technology Facilities Council)

Presentation materials

There are no materials yet.