Bead Test Result of Cavity Beam Position Monitor for PAL-XFEL

10 Sept 2025, 16:00
2h
Teaching Hub 502

Teaching Hub 502

The University of Liverpool 160 Mount Pleasant L3 5TR Liverpool
Poster Presentation MC03: Beam Position Monitors WEP

Speaker

Changbum Kim (Pohang Accelerator Laboratory)

Description

A bead test was conducted to measure the shunt impedance (R/Q) of the cavity beam position monitor (BPM) for PAL-XFEL. R/Q is an important parameter of a cavity BPM because it relates to the signal strength, which determines the resolution of the beam position measurement. In the bead test, a dielectric and metal bead were used to assess the frequency change at various bead positions. The measurement results of the two beads from the monopole cavity were similar, but the results from the dipole cavity differed significantly. This discrepancy can be explained by considering the distribution of the electric and magnetic fields in the cavity structure and the materials of the two beads.