Development of moving long coil magnetic measurement system based on HALF magnets

WEPB092
4 Jun 2025, 16:00
2h
Exhibiton Hall A _Bear (TWTC)

Exhibiton Hall A _Bear

TWTC

Poster Presentation MC7.T09 Normal Conducting Magnets Wednesday Poster Session

Speaker

Weihao Zhu (University of Science and Technology of China)

Description

Hefei Advanced Light Facility (HALF) consists of two main components: the injector and the storage ring, the core of the facility includes nearly one thousand magnets, of which there are more than one hundred conventional dipole magnets and longitudinal gradient bend magnets.The longitudinal integral field and its uniformity of these dipole magnets are typically measured by the Hall probe measurement system, which takes a long time to measure, and the measurement accuracy of the uniformity repeatability of the integral field can only reach 0.01%. The accuracy of the integral field uniformity of the long coil magnetic measurement system can reach less than 0.01%, and the measurement accuracy of the uniformity repeatability of the integral field can reach less than 0.005%. Therefore, to efficiently and accurately measure the uniformity of the integral field, there is a necessity to develop the moving long coil magnetic measurement system. This paper develops a moving long coil magnetic measurement system based on the technical requirements for the measurement of the integral field of dipole magnets in the HALF project.